DISP:MODE
Switches the display between its normal metering mode and a mode in which it displays text sent by the user. The command
uses the character data
Command Syntax
DISPlay[:WINDow]:MODE NORMalITEXT
Parameters
*RST Value
NORM
Examples
DISP:MODE NORM DISPLAY:MODE NORMAL
DISPLAY:WINDOW:MODE TEXT
Query Syntax
DISPlay[:WINDow]:MODE?
Returned Parameters
<CRD> NORMAL or TEXT
Related Commands
DISP DISP:TEXT *RST
DISP:TEXT
Allows character strings to be sent to display. The characters will be displayed when the display mode is TEXT. The LCD
has the following character set:
LCD Character Set
uppercase letters
A through Z (Case-sensitive entry)
digits
0 through 9
punctuation
_ | “ $ < > + - / = ? . : ,
blank space
A display is capable of showing up to 12 characters. However, the three punctuation characters do not count toward the 12-
character limit when they are preceded by an alphanumeric character. When punctuation characters are included, then the
maximum number of characters (alphanumeric + punctuation) that can be displayed is 15. If it exceeds the display capacity,
a message will be truncated to fit and no error message will be generated. If any character in the message is not a member
of the above character set, the character will not be rejected but will be displayed as a "starburst" (all 16 segments of the
character will light).
Command Syntax
DISPlay[:WINDow]:TEXT [:DATA]
Parameters
(See LCD character set)
*RST Value
‘ ‘
Examples
DISP:TEXT "DEFAULT_MODE"
DISPLAY:WINDOW:TEXT:DATA '533.2E-1VOLTS'
Query Syntax
DISPlay[:WINDow]:TEXT?
Returned Parameters
Related Commands
DISP DISP:MODE *RST
Note
IEEE Standard Digital Interface for Programmable Instrumentation requires that a string be