Product specification – UEi Test Instruments KM900PLUS User Manual
Page 29

KANE900 Plus Operators Manual
Page 28
12. PRODUCT
SPECIFICATION
Parameter Resolution
Accuracy Range
Temp Measurement
Flue Temperature
Inlet Temperature
0.1
°C/F
0.1
°C/F
+2.0
°C +0.3% reading
+1
°C +0.3% reading
0-1200
°C/32 -2200°F
with suitable probe
0-50
°C/32-122°F
Gas Measurement
Oxygen
Carbon Monoxide
Carbon Monoxide, H
2
compensated
Nitric Oxide
(optional)
Nitrogen Dioxide
(optional)
Pressure (optional)
Carbon Dioxide
*2
Losses
*2
Efficiency
*2
Excess Air
*2
Temp (Nett)
*2
CO/CO
2
ratio
*2
Poison Index
*2
0.1%
1ppm,mg/m
3
1ppm,mg/m
3
1ppm,mg/m
3
1ppm,mg/m
3
0.1mbar
0.1%
0.1%
0.1%
0.1%
1.0
°C/F
0.0001
0.01%
+0.2%
*1
+20ppm <400ppm
*1
+5% >400ppm
+20ppm <400ppm
*1
+5% <5000ppm
+10% >5000ppm
+5ppm <100ppm
*1
+5% >100ppm
+5% full scale
+5.0% full scale
+0.3% reading
+1.0% reading
+1.0% reading
+0.2%
+2
°C +0.3% reading
+0.0001
+0.01
0-25%
0-4000ppm, 0-5000mg/m
3
0-10,000ppm, 0-12,000mg/m
3
0-5000ppm, 0-6700mg/m
3
100ppm, 200mg/m
3
150 mbar
0-99.9%
0-99.9%
0-99.9%
0-2885.0%
0-1200
°C/32-2200°F
0-0.9999
0-99.99
Pre-programmed Fuels
Natural gas, Town gas, Gascor, Light Oil, Heavy Oil,
Propane, Butane, Anthracite, Coke, Coal, Kinsale Gas.
Dimensions
Weight
Handset
Probe
1kg
220mm x 55mm x 120mm
L240mm x Dia8mm with 285mm long stainless steel shaft,
type K thermocouple and 1.5m long neoprene hose
Ambient Operating Range
-5
°C to +50°C/10% to 90% RH non condensing
Power Supply (battery charger)
Input: 110Vac/220 Vac nominal
Output: 12 Vac off load
Battery Life
>8 hours from full charge
*1
Using dry gases at STP
*2
Calculated