Additional features and operation, Rfprhtemp2000a – MadgeTech RFPRHTemp2000A Wireless Pressure, Humidity and Temperature Data Logger User Manual
Page 5

RFPRHTemp2000A
Additional Features and Operation
Alarm Settings/Rules
In MadgeTech 4, alarms can be created in
the software or programmed directly on
the data logger. Creating alarm rules in the
software allows the user to utilize alarms
during real time recording, and have the
added benefit of features such as e-mail
and SMS notification. An alarm that is
programmed on the data logger will only
alert the user from the device itself, either by
a flashing LED, audible alarm, or local LCD
notification.
Creating a Real Time Alarm Rule
1. Within the Device tab in MadgeTech 4, in the Alarms group, click Manage Rules,
then select New.
2. Enter a name in the Rule name box, and select whether notifications will occur when
All conditions are met or Any condition is met.
3. Select the desired conditions to have met before notifications occur, and select which
notifications will occur from the drop down menus.
4. If selecting to receive notifications by email or text message, click Email settings and
make sure the correct info has been entered. Email and text message notifications will
not be received if settings are incorrect.
5. Click OK. The OK button will be disabled if the Rule name box is empty.
Changing the Alarm Settings of a Device
1. In the Connected devices panel, select the RFPRHTemp2000A device.
2. Click the Properties button under the device tab at the top of the software, and select
the Alarm tab in the Properties window.
3. The High and Low settings indicate reading thresholds at which the alarm becomes active.
4. The Delay setting indicates how long to wait before the alarm becomes active after
passing an alarm threshold.
5. The Use cumulative alarm delay checkbox indicates whether the alarm delay should
reset when reading values fall back within the alarm threshold.
6. Once the alarm is configured appropriately, click Apply to save the settings.
**Please refer to the Help file in the MadgeTech software for more information regarding alarm settings**
3