6
Warning messages and faults
Fault/indication
Possible cause
Remedy
ERROR
OUT OF RANGE
CHANGE BATTERY
No change in
measurement
Does not switch on
Instrument
inaccurate
10% exceeding or under-
running of measuring range
Battery voltage too low
Over pressure applied to
pressuresensor
No power supply
Inaccurate recalibration
Not zeroed
Natural aging of the pressure
sensor
Apply permissible measuring
pressure
Insert new battery
Dispatch instrument for repair
Fit new battery as required
Battery possibly not correctly inserted
Plug in power supply unit correctly
Repeat recalibration
Vent and actuate zero
Carry out recalibration