Dwyer HM28 User Manual

Page 13

background image

13

7

Accessories

Standard 1 9 V block battery

1 operating instructions

Option

Plug-in power supply unit
115V, 60Hz
Leather case with carrying strap
Service set
Hand pump with variobellows 5bar
Handpump 20 bar
SCS test certificate
Adapter RS232 9M-25F
Adapter NPT1/8"
Communication SW package

comprising:

-

RS232-IF cable (9 pole fem.)

-

Communication-software for
MS Windows

-

Measurement places
management software for
MS Windows

6

Warning messages and faults

Fault/indication

Possible cause

Remedy

ERROR
OUT OF RANGE

CHANGE BATTERY

No change in
measurement

Does not switch on

Instrument
inaccurate

10% exceeding or under-
running of measuring range

Battery voltage too low

Over pressure applied to
pressuresensor

No power supply

Inaccurate recalibration

Not zeroed

Natural aging of the pressure
sensor

Apply permissible measuring
pressure

Insert new battery

Dispatch instrument for repair

Fit new battery as required
Battery possibly not correctly inserted
Plug in power supply unit correctly

Repeat recalibration

Vent and actuate zero

Carry out recalibration