Thermo Fisher Scientific ECDOTPII-S Submersible Galvanic Dissolved Oxygen Sensor User Manual
Page 5

www.thermofisher.com
SENSOR STORAGE
If long-term storage of probes is required, empty electrolyte out of probe leaving only a little solution on
the membrane to keep it wet. Leaving a full volume of electrolyte in DO probe long-term without use will
deplete the probes’ anode. See section “Sensor RE-CONDITIONING ” for steps on how to open and close
probe housing.
SENSOR SPECIFICATIONS
Output at 100% saturation
PTFE membrane 24-42 mV
HDPE membrane 36-54 mV
Output at 0% saturation
PTFE membrane <1mV <1mV
HDPE membrane <1mV <1mV
Temperature Range
Max 50
°C
Min 0
°C
Accuracy
+2% when measuring temp equal cal temp.
Response Time
PTFE membrane 2 minutes to reach 90% of final reading
HDPE membrane 1 minute to reach 95% of final reading
Water Flow Rate
Min 2 inch/second across membrane
Wetted Materials
Body
Delrin
Membrane Teflon,
HDPE
Wiring
DO +
Red
DO –
Yellow
TC +
Red, White
TC –
Black
Teflon and Delrin are registered trademarks of E.I. DuPont.