8.9 Setting TDS Factor – P7.0
Program 7 allows the TDS factor to be selected, depending on the application.

.

P .0

d

t s

0

0


P7.1: Setting TDS Factor
1. Key in the password “022” as per procedure in Section 5.
2. Press

▲key until screen displays P7.0 and TDS. Press ENTER.

3. Screen will scroll P7.1, ‘FACT.’ then TDS factor. (Default is 0.70)
4. Press

the

▲or ▼keys to change TDS factor (0.40 to 1.0). Press ENTER to

confirm.

5. Press

▲and ▼keys together (ESCAPE) to return to P7.0.