10 RELAYS

The alpha pH 200 features two SPDT non-powered relays; rated for 6A at 110
VAC, 250 VAC maximum. When your process exceeds the set parameters of a
relay set point, the REL 1 or REL 2 indicator lights up.

To set parameters for relay one and relay two, see Setup programs P1.0 and P2.0.

11 TRANSMITTER FUNCTION

If remote data logging is required, a 4-20 mA current loop can be connected. The
current will be proportional to the pH (or ORP) displayed on the panel display.

pH: 0.00 pH – 4 mA; 14.00 pH = 20 mA
ORP: -999 mV = 4 mA; 1000 mV = 20 mA

The 4-20 mA current loop can drive a load resistance of no more than 200 Ω.

12 ERROR MESSAGES

The following table provides a guideline to enable diagnosis of possible problems
indicated by the messages generated by the meter.

Error Message

Indicates

Possible Cause

Corrective Action

Err1 Calibration

Error

Incorrect buffer solution used

or contamination of buffer

solution resulting to

calibration being attempted

outside the allowable

calibration window

Check if right

buffer solution

was selected or

use fresh buffer

solution.

Err2 Wrong

keypad input

When an invalid key is

pressed in a selected mode.

Select valid

operations

depending on the

selected mode.