Operating Instructions
Thermo Scientific Alpha COND 1000
25
7.7.5
Wash contact (if the relay 3 is set to Wash)
1) Follow directions in 7.7.1 to enter Configuration mode.
2) Press
the
ENTER key until display shows “ALr” or “CLn”.
Choose “CLn” by pressing
▲ or ▼ key. Press the
ENTER key.
3) Press
the
▲ or ▼ key to select the wash cycle (interval
0.1 to 199.9 hours) and press the ENTER key to confirm.
4) Press
the
▲ or ▼ key to select the wash duration (1 to
1999 seconds) and press the ENTER key to confirm.
5) Proceed to 7.7.6, or return to Measurement mode by
pressing the
▲ and ▼ keys simultaneously (escape).
7.7.6
Input Line Resistance Adjust
This function compensates for the line resistance of the cable to its cell.
1) Follow directions in 7.7.1 to enter Configuration mode.
2) Press the ENTER key. Scroll with the
▲ or ▼ keys
until the upper display shows “0.0” and lower display
shows “L.Ad”.
-
L.Ad = line adjuster resistance (0.0 to 100.0)
3) Press
the
▲ or ▼ keys to input the value.
4) Press
the ENTER key to confirm your selection.
5) Proceed to 7.7.7, or return to Measurement mode by pressing the
▲ and ▼
keys simultaneously (escape).
7.7.7
Reverting to factory default settings
Use this parameter to reset all settings to factory default.
Changing from “no” to “YES” and pressing the ENTER
key resets all settings to factory default.
WARNING: If you select yes, all the settings you have
made will be overwritten as a result!
1) Follow directions in 7.7.1 to enter Configuration mode.
2) Press
the ENTER key. Scroll with the
▲ or ▼ keys
until the upper display shows “no” or “YES”, and the lower display shows “deF”
(default).
3) Press
the
▲ or ▼ keys to select no or yes. Selecting “YES” and pressing the
ENTER key will overwrite all setting you have made and automatically return
you to Measurement mode.
4) Press
the ENTER key to confirm your selection and to return to Advanced Set-
up mode, or return to Measurement mode by pressing the
▲ and ▼ keys
simultaneously (escape).