Thermo Fisher Scientific CyberScan 6000 Series Meters v.2 User Manual
Thermo Fisher Scientific Tools

86
Conductivity Setup: Select Reference Temperature
PC6000, CON6000, PC6500 and PCD6500 meters
Technology Made Easy ...
Set Default Temperature
Conductivity is a temperature dependent measurement. This option allows you to select a default
temperature that the meter will use along with the temperature coefficient to calculate a
temperature corrected conductivity measurement.
If an ATC probe is used any value entered as a default temperature will be overridden in favor of
the actual measured temperature of the solution provided to the meter by the ATC probe.
The default temperature can be set from -10°C to 110°C.
To set Default Temperature
1. Select temperature units by touching the appropriate unit button: C (Celsius), F
(Fahrenheit) or K (Kelvin).
2. Touch the Default Temperature box and use the numeric keypad to enter the desired
default temperature (-10°C to 110°C).
3. Press Enter in the keypad to return to conductivity Setup screen.
Set Reference Temperature
This setup option allows you to select the reference temperature to be used in the determination
of the conductivity of your sample. The reference temperature is the temperature to which the
conductivity readings would be normalized.
The reference temperature can be set from 15°C to 30°C.
To set Reference Temperature
1. Select temperature units by touching the appropriate unit button: C (Celsius), F
(Fahrenheit) or K (Kelvin).
2. Touch the Reference Temperature box and use the numeric keypad to enter the
desired default temperature (15°C to 30°C).
3. Press Enter in the keypad to return to conductivity Setup screen.
Conductivity cells provided have a built-in Temperature sensor.
Note: The units of all temperature parameters must match. The meter will automatically
modify the temperature units of Default Temperature and Reference Temperature to match
the last choice made on any of these parameters.
Document Outline
- Getting Started
- Using the 6000 Series Meter
- Getting Started
- Using the 6000 Series Meter
- Conductivity Setup
- Getting Started
- Using the 6000 Series Meter
- Ion Setup
- Getting Started
- Using the 6000 Series Meter
- Conductivity Setup
- Resistivity Setup
- Getting Started
- Using the 6000 Series Meter
- DO Setup
- DO Setup Screen and Function Buttons 63
- Access DO Setup 64
- Set Sample ID# 65
- Select Measurement Units 66
- Select Calibration Mode 66
- Select Stable Indicator 67
- Select Auto Read Mode 67
- Select Temperature Units 68
- Set Barometer 68
- Set Salinity 69
- Set Alarm Limits 69
- Set Print Criteria 70
- Set Data Storage Criteria 71
- Set Display Criteria Setup 72
- View Stored Data 73-74
- BOD Setup
- BOD Setup Screen and Function Buttons 75
- BOD Configuration 76
- OUR Setup
- OUR Setup Screen and Function Buttons 77
- OUR Configuration 78
- SOUR Setup
- SOUR Setup Screen And Function Buttons 79
- SOUR Configuration 80
- Getting Started
- Using the 6000 Series Meter
- Ion Setup
- Getting Started
- Using the 6000 Series Meter
- Ion Setup
- TDS (Total Dissolved Solids) Setup
- Blk 55, Ayer Rajah Crescent,
- Web-site: www.eutechinst.com