Low Slope or

standards contaminated

prepare fresh standards

No Slope

or incorrectly made


standard used as ISA

use ISA


ISA not used

use recommended ISA


membrane dirty

polish membrane; use

or oxidized

methanol-formaldehyde solution


air bubble on

remove bubble by

membrane

re-dipping probe



"Incorrect Answer"

incorrect scaling

plot millivolts on the linear axis. On

(but calibration

of semi-log paper

the log axis, be sure concentration

curve is good)

numbers within each decade are

increasing with increasing

concentration


incorrect sign

be sure to note sign

of millivolt reading correctly


incorrect standards

prepare fresh standards


wrong units used

apply correct conversion factor:

10

-3

M = 207 ppm Pb

+2

= 96 ppm SO

4

-2


complexing agents

check section entitled

in sample

Precipitation and Complexation

;

use titration method