beautypg.com

Thermo Fisher Scientific Ion Selective Electrodes Iodide User Manual

Page 9

background image

Instruction Manual

Iodide Electrode

9

TABLE 2: Maximum Allowable Ratio of Interfering Ion to Iodide Ion

Interference

Maximum Ratio

Cl

-1

(1)

1.0X10

6

S

2

O

3

-2

(2)

1.0X10

5

Br

-1

(1)

5.0X10

3

CN

-1

(3)

4.0X10

-1

S

-2

(3)

1.0X10

-6

(1) Gran's plot titration can be used to measure mixed halides in solution.

(2) These substances represent complexing species whose maximum level can be exceeded

without electrode damage. Values shown represents a 1% error.


(3) Add Ni

+2

to remove sulfide or cyanide interferences.

As an example of Table 2's use when using the iodide ion electrode:

What is the maximum level of iodide tolerable in a sample whose iodide concentration is 10-4M?

Using Table 2, the maximum ratio is:

[Br

-1

]

⎯⎯⎯⎯ = 5X10

3

[I

-1

]


or [Br

-1

] = 5X10

3

[I

-1

]

= 5X10

3

(1X10

-4

)


[Br

-1

] = 5X10

-1

M maximum bromide concentration for no interference

Complexation

Total concentration (C

t

) consists of free ions (C

f

) and complexed or bound ions (C

c

) in solutions:

C

t

= C

f

+ C

c


Since the electrode only responds to free ions, any complexing agent in the solution reduces the
measured concentration of ions.

Iodide ions complex with some metal ions. Table 3 lists the levels of complexing metals causing a
20% error.

TABLE 3: Levels of Complexing Agents Causing a 20% Error at 1.0X10

-4

M Iodide


Ion Concentration

Bi

+3

2.0X10

-5

M ( 4 ppm)

Cd

+2

5.0X10

-4

M ( 50 ppm)

Pb

+2

5.0X10

-3

M (1,000 ppm)