Cyanide Electrode

17

Low Slope or

standards contaminated

prepare fresh standards

No Slope

or incorrectly made


ISA not used

use recommended ISA


standard used as ISA

use ISA


electrode exposed

check section entitled

to complexing agents

Complexation


air bubble on membrane

remove bubble by

re-dipping probe



"Incorrect Answer" incorrect scaling of

plot millivolts on the linear

(but calibration

semi-log paper

axis. On the log axis, be sure

curve is good)

concentration numbers within

each

decade

are

increasing

easing

concentration.


incorrect sign

be sure to note sign

of millivolt number correctly


incorrect standards

prepare fresh standards


wrong units used

apply correct conversion factor:

10

-3

M = 26.0 ppm CN-


complexing agents

check section entitled

in sample

Complexation