Thermo Fisher Scientific Ion Selective Electrodes Cyanide User Manual
Page 16
Cyanide Electrode
Instruction Manual
16
TROUBLESHOOTING HINTS
Symptom Possible Causes
Next Step
Out of Range defective meter
check meter with shorting
Reading
strap (see meter
instruction manual)
defective electrode
check electrode operation
electrodes not
unplug electrodes and reset
plugged in properly
reference electrode
be sure reference electrode
not filled
is filled
air bubble on membrane
remove bubble by
re-dipping electrode
electrodes not in solution
put electrodes in solution
Noisy or Unstable defective meter
check meter with shorting
Readings (readings
strap
continuously or
rapidly changing) air bubble on membrane
remove bubble by
re-dipping electrode
electrode exposed
soak electrode in
to interferences
cyanide standard
ISA not used
use recommended ISA
meter or stirrer not
ground meter or stirrer
grounded
defective electrode
replace electrode
Drift (reading
samples and standards
allow solutions to come to
slowly changing at different
room temperature before
in one direction) temperatures
measurement
electrode exposed
check section entitled
to complexing agents
Complexation
incorrect reference
use recommended
filling solution
filling solution