6. PROBE CARE AND MAINTENANCE

For best results, always clean the temperature probe with clean tissue paper after
measurement to keep the probe in good working condition. Wash the probe
thoroughly with tap water if necessary to wash off any residue. Should there be any
thin oil film sticking on the probe, wash with mild detergent or soap and warm
water. Rinse probe thoroughly under running water. Blot it dry and clean off with
clean tissue.

To remove the probe, simply hold firmly onto the probe’s plastic holder and pull the
connector out of the meter’s socket. Store both the probe and meter into its original
packaging when not in use.

7. TROUBLESHOOTING

Problem Cause

Solution

Power on but no
display

a) Batteries not in

place.

a) Insert batteries.

b) Re-insert batteries in correct

polarity.

“oPEn” display on
LCD

a) Probe not connected

a) Make sure probe is firmly

connected.

“Ur” or “Or” display
on LCD

a) Measurement over
(Or) or Under (Ur)
range

a) Ensure temperature taken is

within meter’s specification.

“LO” annunicator in
the LCD

a) Low battery

a) Replace batteries with fresh

ones.

Unstable reading

a) Electrode not deep

enough in sample

b) Dirty probe

c) Broken probe

d) External “noises” or

induction caused by
nearby electrical
motor

a) Place probe deeper in

sample.

b) Clean probe.

c) Replace electrode.

d) Remove or switch off

interfering motor.

Slow response

a) Dirty probe

a) Clean probe