Probe care and maintenance, Troubleshooting – Thermo Fisher Scientific EcoScan Temp JKT User Manual
Page 14

11
6. PROBE CARE AND MAINTENANCE
For best results, always clean the temperature probe with clean tissue paper after
measurement to keep the probe in good working condition. Wash the probe
thoroughly with tap water if necessary to wash off any residue. Should there be any
thin oil film sticking on the probe, wash with mild detergent or soap and warm
water. Rinse probe thoroughly under running water. Blot it dry and clean off with
clean tissue.
To remove the probe, simply hold firmly onto the probe’s plastic holder and pull the
connector out of the meter’s socket. Store both the probe and meter into its original
packaging when not in use.
7. TROUBLESHOOTING
Problem Cause
Solution
Power on but no
display
a) Batteries not in
place.
a) Insert batteries.
b) Re-insert batteries in correct
polarity.
“oPEn” display on
LCD
a) Probe not connected
a) Make sure probe is firmly
connected.
“Ur” or “Or” display
on LCD
a) Measurement over
(Or) or Under (Ur)
range
a) Ensure temperature taken is
within meter’s specification.
“LO” annunicator in
the LCD
a) Low battery
a) Replace batteries with fresh
ones.
Unstable reading
a) Electrode not deep
enough in sample
b) Dirty probe
c) Broken probe
d) External “noises” or
induction caused by
nearby electrical
motor
a) Place probe deeper in
sample.
b) Clean probe.
c) Replace electrode.
d) Remove or switch off
interfering motor.
Slow response
a) Dirty probe
a) Clean probe