16. CALIBRATION SCREEN

The thermometer is factory calibrated and

does not require calibration before use.

The Calibration function allows single

point calibration of the thermometer, at

0°C (32°F) to compensate for

thermocouple off-set error. It is NOT

necessary to perform a field calibration to

obtain specified meter accuracy. Use the

field calibration feature to improve

thermometer/probe accuracy or to

compensate for thermocouple drift...

Before go into the calibration mode, must

enter the password. Press F2 to change to

the next digit. (Default Password is 9900)

There are three calibration options:

Offset – adjusts at a single point

Slope – adjusts at two points

Match – adjusts readings on T1 to match

those on T2. Or adjusts T2 to match T1.