Techne TU-20D User Manual
Page 26

26
spEcIfIcATIOn
TU-20D
Operating temperature range
-40 to +200°C
Temperature range without cooling mechanism
Ambient +5 to 200°C
Temperature selection
Digital
Temperature display
Digital LED
Temperature stability
±0.005°C
Set point accuracy
±1°C
Method of control
PID
Temperature sensor
PRT
Nominal heater power
230V
1800W
120V
1500W
100V
1250W
Maximum watts density
6.2 W/cm
2
Pump capacity
Maximum flow
10 l/min
Maximum pressure
145 mbar
Communications
RS232
The TU-20D has passed
EN 50081-1:1992 Electromagnetic Compatibility; Generic emission standard.
EN 50082-1:1992 Electromagnetic Compatibility; Generic immunity standard (Performance criterion B).
The specification was achieved in an 8 litre bath with a ball blanket according to DIN 58966.
Protection against hazards
IP30
Safety device cassification
2