CLEANING THE SENSOR

5.1

Wipe the electrodes end with a soft cloth and rinse with clean water.

5.2

Clean the sensor electrodes in a soap solution using a soft cloth, brush, Q-tip or pipe
cleaner. If soap solution cleaning is unsuccessful, place the sensor electrode in 10% HCl
(hydrochloric acid) for 1 to 5 minutes to remove the contaminants that are acid soluble. Acid
clean only if absolutely necessary. Never use abrasive cleaners on the sensor or its
electrodes.

6.0

TROUBLESHOOTING AND SERVICE

6.1

Use the troubleshooting section in the analyzer instruction manual to determine whether the
sensor or analyzer is defective.

6.2

Should service, parts, assistance in troubleshooting or repairs be required, please contact
your AquaMetrix representative or the factory directly:

AquaMetrix Inc.
1245 Maple Hill Court, Unit 7
Newmarket, Ontario
L3Y 9E8 Canada

Tel: (800) 742-1413

(905) 954-0841

Fax: (905) 954-0415

Email: [email protected]

An RMA# is required for all returning items. Please contact AquaMetrix if you wish to send a
product for repair or evaluation. A description of the malfunction as well as the proper return
address should accompany all sensors returning for repair.