SCALING

Two resistors Ra and Rb may be
used to alter the full scale reading
(FSR) of the meter - see table. Cut
the PCB track across Ra before
fitting Ra. The meter will have to
be re-calibrated by adjusting the
calibration potentiometer on the
rear of the module.

+

-

Ra

Rb

IHI

Vin or in

I

ILO

EM32-1B-LED

FSR

Ra

Rb

2V

910k

100k

20V

1M

10k

Vin

200V

1M

1k

2000V*

1M

100R

200 A

0R

1k

2mA

0R

100R

Iin

20mA

0R

10R

200mA

0R

1R

m

Voltage

Current

*Ensure that Ra is rated for high voltage use.

Must use external scaling resistors (Do not cut track across Ra).

APPLICATIONS

Do not connect more than one meter to the same power supply if the meters cannot use the same signal
ground. Taking any input beyond the power supply rails will damage the meter.

EM32-1B-LED

Round Hole Mounting LED Voltmeter

IHI

0V

0V

+

V+

Measuring a single ended input
voltage referenced to supply, i.e. the
input voltage and the meter's power
supply share the same 0V rail.
Ensure Jumper link J1 is not fitted.
Ensure Jumper link J2 is fitted.

IHI

ILO

0V

0V

+

-

±200mV

+4.75 to +5.25V

+4.75 to +5.25V

V+

Measuring an input voltage referenced
to a floating supply, i.e. the input
voltage and the meter's power supply
are isolated from each other.
Ensure Jumper link J1 is fitted.
Ensure Jumper link J2 is not fitted.

+4.75 to +5.25V

IHI

ILO

0V

0V

R

V+

Measuring current.
The meter's supply is isolated
from the current being measured.
Ensure Jumper link J1 is fitted.
Ensure Jumper link J2 is not fitted.

R = 0.2

I

FSR

RB

RA

Cut the PCB track across Ra
before fitting Ra.