beautypg.com

Custody transfer / flow calibration – Martel Electronics BetaGauge 321A-Ex User Manual

Page 26

background image

Setup 4: The lower display is set to [P1] switch test, the other displays are off.

Setup 5: The upper display is set to [P1], the middle display is set to [P2] and the lower
display is set to RTD.

10. Custody Transfer / Flow Calibration

Warning

Check entity parameters before making any connections to this device.

The Model 311A-Ex/321A-Ex is ideal for flow computer calibration. Every manufacturer of
flow computers has a different calibration procedure, but most call for calibration of three
parameters: static pressure, differential pressure and temperature. To facilitate these
measurements recall setup #5 on the 311A-Ex/321A-Ex.

Note: The pressures in the UPPER, and MIDDLE displays can be changed to [P1],
[P2], and EXT.

1.

Connect the calibrator to your static and differential pressures. ([P1], [P2]) Then
connect the RTD sensor to the calibrator.

2. Using the reading of your RTD, static, and differential pressures make sure the flow

computer has the correct reading. If not, adjust the flow computer as necessary.

24

This manual is related to the following products: