10
̶
6. OPERATING INSTRUCTIONS
6-1 Current Measurements
WARNING
●
Do not make measurements where the potential is greater
than 600 V AC. This may cause shock hazard and damage
to the instrument or equipment under test.
●
The Transformer Jaws are made of metal and their tips
are not insulated. Be especially careful about the hazard
of possible shorting where the equipment under test has
exposed metal parts.
●
Do not open the battery compartment cover when making
measurements.
CAUTION
●
Take sufficient care to avoid shock, vibration or excessive
force when handling the instrument. Otherwise, precisely
adjusted Transformer Jaws will be damaged.
●
When Transformer Jaws do not fully close, never try to
close them by force, but make them free to move and
try again. If a foreign substance is stuck in the jaw tips,
remove it. If the jaw tips have been deformed, correct so
that each tip is properly aligned. Otherwise, the jaws will
be damaged and warranty may not cover the repair cost.