KYORITSU 2009A User Manual
Page 9

― 7 ―
●Operating System
Dual Integration
●Display
Liquid crystal display with a maximum
count of 4000 plus annunciators
●Overrange Indication
"OL" is shown on the display
●Response Time
Approx. 2 seconds.
●Sample Rate
About 3 times per second.
●Location for use
Indoor use, Altitude up to 2000m
●Temperature and Humidity
23 +/-5℃, relative humidity up to 75%
for Guranteed Accuracy without condensation
●Operating Temperature
0-40℃, relative humidity up to 85%
and Humidity
without condensation
●Storage Temperature
-20-60℃, relative humidity up to 85%
and Humidity
without condensation
●Power Source
6F22(DC9V) or equivalent battery
●Current Consumption
Approx. 15mA max.
●Sleep function
Automatically powered down in about 30
minutes after the last switch operation
(power consumption: about 200μA)
●Overload Protection
DC/AC current ranges: 2400A AC for 10sec
DC/AC voltage ranges: 1200V AC/DC for 10sec
Resistance range: 600V AC/DC for 10sec
●Withstand Voltage
5500V AC, 50/60Hz for 1 minute
between electrical circuit and housing
case or metal part of the jaws
●Insulation Resistance
10MΩ or greater at 1000V between
electrical circuit and housing case or
metal part of the jaws
●Conductor Size
Approx. 55㎜ diameter max.
●Dimensions
250(L)x105(W)x49(D)mm
●Weight
About 530g
●Accessories
Test leads M-7017
6F22 battery
Carrying case M-9094
Instruction manual
Recorder output Plug M-8201
●Optional Accessories
Multi-Tran M-8008
Recorder M-5100A, etc.
Output Lead M-7014.