KYORITSU 6011A User Manual
Page 9
― 7 ―
Function
Rated
Voltage
Measuring Range
Accuracy
Voltage Measurement-
100-250V
100-250V
3% rdg
To prevent wrong connection of test leads and to maintain safety, the dedicated
terminals used for continuity and insulation tests are automatically covered
when using the terminals for loop impedance, PSC and RCD tests
● Typical Number of Tests (central tendency for supply voltage up to 8V at R6)
Insulation Resistance Ranges : Approx. 500 times min. at load 0.5MΩ
Continuity Ranges
: Approx. 300 times min. at load 1Ω
LOOP.PSC.RCD Ranges
: Operational lifetime:5h (In case of
continuous duty)
Operating error
● Operating Errors (IEC 61557-2,-4)
Function
Range
250V
500V
1000V
0.25~199.9MΩ
0.50~199.9MΩ
1.00~199.9MΩ
0.20 ~ 19.99Ω
10.0 ~ 199.9Ω
100 ~ 1999Ω
20Ω
200Ω
2000Ω
Measuring range to keep
operating error
Maximum percentage
operating error
±30%
Insulation
Continuity
Function
RCD X 1/2
Accuracy
Rated Voltage
(AC)
Trip Current
Trip Current
Duration
230V+10%-15%
50Hz
230V+10%-15%
50Hz
230V+10%-15%
50Hz
RCD X 1
RCD X 5
10/30/100/300/
500/1000 mA
10/30/100/300/
500/1000 mA
30/100/300mA
(Note:on × 5
range maximum
current that can
be generated is
1A)
2000ms
2000ms
1000mA@200ms
50ms
Trip Current:
-10% +0% of
range at 230V
Trip Current:
+10% -0% of
range at 230V
Trip Current:
± 10% of
range at 230V
10 mA
Trip Current:
+10% -0% of
range at 230V
Trip Time
±(1% rdg
+ 3dgt)