3. SPECIFICATION
Measurement Specification
Function
Continuity
Open Circuit
Voltage (DC)
Short Circuit
Current
Range
Accuracy
Greater than 6 V
Greater than
200mA
20/200/2000Ω
Auto- Ranging
± (1.5%rdg + 3dgt)
Function
Open Circuit
Voltage (DC)
Rated Current
Range
Accuracy
Insulation
Resistance
250V+40%−0%
500V+30%−0%
1000V +20%−0%
1mA or greater
@ 250kΩ
1mA or greater
@ 500kΩ
1mA or greater
@ 1MΩ
20/200 MΩ
Auto-Ranging
20/200 MΩ
Auto-Ranging
20/200 MΩ
Auto-Ranging
± (1.5%rdg + 3dgt)
± (1.5%rdg + 3dgt)
± (1.5%rdg + 3dgt)
Accuracy
± (3%rdg + 4dgt)
± (3%rdg + 8dgt)
± (3%rdg + 4dgt)
Range
20Ω
200Ω
2000Ω
Nominal Test
Current at 0
Ω
External Loop
25A
15mA
15mA
Function
Rated Voltage
(AC)
Loop
Impedance
230V+10%-15% 50Hz
230V+10%-15% 50Hz
230V+10%-15% 50Hz
Prospective
Short Circuit
Current (PSC)
230V+10%-15%
50Hz
15mA
25A
25A
200A
2000A
20kA
PSC accuracy is a
derived from measured
loop impedance
specification and
measured voltage
specification
@ KAMP10 Test lead
Accuracy of this instrument will be affected on the continuity
range if used in proximity of radio transmitters