beautypg.com

Specification – KYORITSU 4140 User Manual

Page 10

background image

̶ 8 ̶

5. Specification


5.1 Measurement Specification

LoopImpedance

Function

(Operating Voltage)

Rated Voltage

Range

(Auto-Ranging)

NominalTestCurrent

at0ΩExternalLoop:

Magnitude/Duration

(*1)

Accuracy

Guaranteed Voltage

Range

L-PE

ATT OFF

(100∼

280V)

(45∼65 Hz)

230V(50/60Hz)

L-PELOOP:

20Ω:0.00-19.99Ω

200Ω:20.0-199.9Ω

2000Ω:200-1999Ω

PFC/PSC:
2000A:0-1999A
20kA:2.00-19.99kA

L-PE:

20Ω:
6A/20ms

200Ω:
2.3A/20ms

2000Ω:
15mA/250ms

L-N:
6A/20ms

±(3%rdg+4dgt)

(*2)

230 V (+10%/-15%)
(50/60 Hz)±1%

ATT ON

(100∼

280V)

(45∼65 Hz)

230V(50/60Hz)

L-PELOOP:

20Ω:0.00-19.99Ω

200Ω:20.0-199.9Ω

2000Ω:200-1999Ω

PFC/PSC:
2000A:0-1999 A
20kA:2.00-19.99 kA

(L-N<20Ω)

L-N:6A/60ms

N-PE:10mA
/approx. 5s

±(3%rdg+6dgt)

(*2)

230 V (+10%/-15%)
(50/60 Hz)±1%

L-N/L-L

(100∼500V)

(45∼65 Hz)

L-N:230V(50/60Hz)

L-L:400V(50/60Hz) L-N/L-LLOOP:

20Ω:0.00-19.99Ω

PSC:
2000A:0-1999 A
20kA:2.00-19.99 kA

20Ω:6A/20ms

L-N:

±(3%rdg+4dgt)

L-L:

±(3%rdg+8dgt)

(*3)

L-N:
230 V (+10%/-15%)

L-L:

400V(+10%/-15%)
(50/60 Hz)±1%

*1: at 230V

*2:Accuracy of L-N LOOP displayed on the Sub Display is

synchronizedwiththeoneatL-N/L-Lfunction.

  PSC/PFCAccuracyisderivedfrommeasuredloopimpedance

specification and measured voltage specification.

*3:PSCAccuracy is derived from measured loop impedance

specification and measured voltage specification.