beautypg.com

KYORITSU 2033 User Manual

Page 5

background image

5. Preparation for Measurement

5−1

Checking Battery Voltage

Set the Function Selector switch to any position other than "OFF".
When the display is clear without

showing, proceed to

measurement.
When the display blanks or

is shown, replace the batteries

according to section 8: Battery Replacement.

NOTE

The Sleep function automatically turns the instrument off in
about five minutes after the last switch operation. Therefore,
the display may be blank with the Function Selector switch set
to a position other than "OFF".
To operate the instrument in this case, set the switch back to
the OFF position, then to the desired position, or press any
button. If the display still blanks, the batteries have
exhausted. Then, replace the batteries.

5−2

Checking Switch Setting

Make sure that the Function Selector switch is set to the correct
position and the Data Hold function is deactivated. Otherwise,
desired measurement cannot be made.

−6−

6. Measurement

6−1

AC Current Measurement

DANGER

●Do not make measurement on a circuit above 300V AC. This

may cause shock hazard.

●Do not make measurement with the battery compartment

cover removed from the instrument.

(1) Set the Function Selector switch to the "

A" position. "AC"

should be shown on the lower left corner of the display.

(2) Press the jaw trigger to open the transformer jaws and clamp them

onto the conductor under test, then take the reading on the
display. Position the conductor at the center of the transformer
jaws for accurate measurement.

Note :
◇During current measurement, keep the transformer jaws fully

closed. Otherwise,accurate measurement cannot be made. The
maximum measurable conductor size is approx. 24mm in diameter.

◇Unlike in DC current measurement, zero adjustment is not

necessary in AC current measurement. There is no polarity in the
reading either.

−7−

Correct

Wrong