Instrument layout – KYORITSU 1009 User Manual
Page 6
― 16 ―
● Operating Environmental
−indoor use
conditions
−altitude up to 2000m
● Temperature & Humidity range
(guaranteed accuracy)
23℃±5℃
Relative humidity: less than 75%
● Operating Temperature &
Humidity range
0℃∼+40℃
Relative humidity: less than 80%
● Storage Temperature &
Humidity range
-20℃∼+60℃ Relative humidity: less than 70%
● Insulation Resistance
It should be more than 10MΩ/DC1000V between electrical circuit and enclosures.
● Withstand Voltage
It should be more than AC3700V/ for one minute between electrical circuit and enclosures.
● Overload Protection
Voltage function : 400mV range
:
250V(RMS.) 10seconds
except for 400mV range
:
600V(RMS.) 10seconds
Resistance function
:
250V (RMS.) 10seconds
Capacity function
:
250V (RMS.) 10seconds
Frequency function
:
250V (RMS.) 10seconds
Current function : uA, mA
:
Protected by 600V0.5A fuse
A
:
Protected by 600V/10A fuse
● Dimensions / Weight
approx. 155(L) x 75(W) x 33(D) mm / approx. 260g(including batteries)
● Power source
Two R6P(AA)1.5V or equivalent
● Accessories
Test lead 1set / R6P (AA) 2p.c.s / Holster 1 /Instruction manual 1
● Fuse
F 600V/500mA(fast acting type), φ6.3 x 32mm / F 600V/10A(fast acting type), φ6.3 x 32mm
CAUTION
The voltage shown above is the overload protection (overvoltage protection) for the instrument.
Make sure not to exceed the value of voltage shown above.
4. Instrument Layout
①Function Selector Switch
②Display
③RESET Key
④SELECT Key
⑤RANGE Key
⑥REL Key
⑦HOLD Key
⑧Hz/DUTY Key
⑨Measuring terminal (V/Ω/Hz)
⑩Measuring terminal (COM)
⑪Measuring terminal (mA)
⑫Measuring terminal (A)
⑬Test Leads
⑭Holster