Operating Instructions DWM 2000 D

2.3.3 Summary of programming menus

Fct. n°

Text

Description and settings

1.0.0.

TEST

Main menu 1.0.0.

1.1.0.

CHECK ALL

Sub menu 1.1.0. for check of electronic components

1.1.1.

MAG. FEQ.

Frequency of magnetic field
10 Hz

≤ frequency ≤ 14.5 Hz, operating mode

1.1.2.

FULL SCALE

Programmed full scale
1 m/s

≤ full scale ≤ 8 m/s

1.1.3.

U REF

Internal voltage reference
U Ref. = 2.5 V

1.1.4.

AMPLI

Test value of amplifier control loop
Value

≥ 40 in test mode

1.1.5. EEP

CHECKS

Result

of data check in EEPROM

“XXX”: measured value = test has been successful
“XXX + ALARM”: test has failed and alarm mode has been
activated (current output value < 4 mA).

1.1.6. EP

CHECKS

Result

of data check in EEPROM

Value = 22309

1.1.7.

ZERO KEY

Result of zero calibration push button test
“OK”: test has been successful; “ALARM”: The test has failed
and alarm mode has been activated (current output < 4 mA).
This is only a test and does not reset the instrument to zero.
Refer to the installation manual to use the Zero Cal button on
the DWM 2000 electronics block correctly.

1.2.0.

DIAGNOSTIC

Sub menu 1.2.0. Diagnostic

1.2.1

ALARM MODE

Deactivation of the alarm mode (current output < 4 mA) caused
by test failure (see menu 1.1.2. to 1.1.7).
“YES”: alarm mode is enabled
“NO”: alarm mode is disabled

1.2.2.

REG FAIL.

All the error messages that have occurred since the first power-
up are listed. Maximum storage capacity: 32 messages
See section “2.5 Error Message List” for the meaning of error
messages.

1.2.3. TEST

MODE

Activation of PRODUCT (factory auto-diagnostic test mode) or
CUSTOM (field auto-diagnosis test mode = less severe).
Default setting: CUSTOM (field test mode)

1.2.4. ALARM

COUNT

Reset

of the alarms counter

2.0.0. SET

UP

PARAMETER

Main menu 2.0.0.

2.1.0.

DATA BASE

Sub menu 2.1.0. Base data

2.1.1.

FULL SCALE

Not available.

2.1.3. GK

VALUE

Primary

head calibration constant 0.8

≤ G

K

≤ 1.300

See section “1.1.1. Flow calibration” on how to recalculate G

K

.

2.1.4.

CORRECTION

Activation of the low flow linearization for velocity below 3 m/s.
Select “YES” or “NO”, default setting:“YES”