Operating Instructions DWM 2000 D

2.4

Parameters stored in the EEPROM (menu 2.4.3)


Parameter Comment

Typical

value

CHECKS 1

Check EEPROM n°1

-

CHECKS 2

Check EEPROM n°2

-

CHECKS BYTE 1

Check EEPROM

-

CHECKS BYTE 2

Check EEPROM

-

CPT ALARM.

Counting of all the error messages
since the first power-up.

-

CPT ALARM.2:

Counting of all the error messages
since the last reset.

-

TEST

Indication of auto diagnostic test
level

CUSTOM

CORRECTION
YES/NO

Indication of activation of low velocity
linearization.

YES

EP CHECKS

Result of the data check in EPROM
(see menu 1.1.6.)

Value = 22309
With soft V1.02

FS

Display of the programmed full scale
values (see menu 1.1.2.)

1 m/s ... 8 m/s

GK

Primary head calibration constant
(see menu 2.1.3.)

0.8 ... 1.3

U REF

Internal voltage reference
(see menu 1.1.3.)

2.4. ... 2.6.

T CST

Time constant (see menu 2.2.4.)

5 s

TEST AMP

Test value of the amplifier control
loop (see menu 1.1.4.)

75 ... 95

FM

Frequency of the magnetic field
(see menu 1.1.1.)

10 ... 14.5 Hz

DIAMETER

Diameter of pipeline
(see menu 2.4.6)

≥ 50 mm