Error message list


Listed below are the messages which can appear in menu 1.2.2. This function stores all the faults
that have occurred since the first connection to power.


Error
messages

Comment

Actions

MAG. FREQ

No magnetic field frequency out of range
10 Hz

≤ frequency ≤ 14.5 Hz

Replace the electronics module

AMPLI

Dysfunction of the amplifier loop

Replace the electronics module

F.S.

Programmed full scale out of range
(>8 m/s or <1 m/s).

Replace the electronics module

ZERO KEY

Dysfunction of ZERO key (short circuit)

Replace the electronics module

EP CHECK

Data loss in EPROM (software)

Replace the electronics module

EEP CHECK

Data loss in EEPROM (calibration and
calculation data)

Replace the electronics module

U REF

Amplifier voltage reference is damaged

Replace the electronics module

CURR. OUT

Incorrect position of the internal current output
switch

Replace the electronics module

FS SWIT EEP

Modification of the full scale power on

Program the full scale power off

ZERO

Velocity measurement during the zero
adjustment is more than 0.2 m/s.

Deactivate the alarm mode and
adjust the zero again