2.7 Electrical data

Connect the sensor only to intrinsically-safe certified circuits. Attend the following maximum
values for the sensor when connecting:

• U

i

= 30 V

• I

i

= 120 mA

• P

i

= 1.0 W

• L

i

= 10 µH

• C

i

= 5 nF

The internal electrode circuit is limited to the following maximum values:

• U

o

= 7.8 V

• I

o

= 10 mA

• P

o

= 20 mW

CAUTION!
Capacity and inductance of the connecting cable have to be considered.

INFORMATION!
For information about nominal data and the HART interface consult the standard documentation.

.book Page 9 Friday, July 11, 2014 2:56 PM