INSTALLATION

3

13

OPTIBAR P 2010 C

www.krohne.com

04/2014 - 4002622502 - MA OPTIBAR P2010C R02 en

3.3.2 Installation steps for connections acc. to EN 837

• When sealing, use a suitable gasket that corresponds to the product and pressure to be

measured (e.g. a copper gasket).

• Ensure that the raised face of the receiving part has a smooth surface.
• Screw the device into the thread by hand.
• Then tighten the device with the wrench.

3.3.3 Installation steps for NPT connections

• When sealing, use a suitable sealing agent that is compatible with the product (e.g. PTFE

tape).

• Screw the device into the thread by hand.
• Then tighten the device with the wrench.

Wrench size

Tightening torque [Nm]

G¼"

approx. 20

G½"

approx. 50

Wrench size

Tightening torque [Nm]

¼

NPT

approx. 30

½

NPT

approx. 70