CONNECTION DIAGRAMS

3

13

TIDALFLUX 2300 F

www.krohne.com

6/2013 - 7312442200 - AD EX TIDALFLUX 2300 - en - R02

3.2 Signal cable B

The signal cable B is a triple screen shielding cable, according to EN/IEC 60079-14 clause 12.2
(intrinsic safety).

Figure 3-2: Construction of signal cable B

1 Stranded drain wire for the inner shield (10), 1.0 mm

2

Cu / AWG 17 (not insulated, bare)

2 Insulated wire (2), 0.5 mm

2

Cu / AWG 20 with stranded drain wire (20) of shield

3 Insulated wire (3), 0.5 mm

2

Cu / AWG 20 with stranded drain wire (30) of shield

4 Outer sheath
5 Insulation layers
6 Stranded drain wire (6) for the outer shield (60), 0.5 mm

2

Cu / AWG 20 (not insulated, bare)