TECHNICAL DATA

28

OPTIFLUX 6000

www.krohne.com

11/2013 - 4001723703 - HB OPTIFLUX 6000 R03 en

Materials

Sensor housing

DN2.5...15: Stainless steel Duplex (1.4462)

DN25...150: Stainless steel AISI 304 (1.4301)

Measuring tube

Stainless steel AISI 304 (1.4301)

Adapters

Stainless steel AISI 316 L (1.4404)

Other materials on request.

Liner

PFA

Connection box

(F-version only)

Standard:

Standard:

Standard:

Standard:

Aluminum, Polyurethane coated

Option:

Option:

Option:

Option:

Stainless steel AISI (1.4408)

Electrodes

Standard:

Standard:

Standard:

Standard:

Hastelloy

®

C

Option:

Option:

Option:

Option:

Hastelloy

®

B2, platinum, stainless steel, tantalum, titanium

Gaskets

Standard:

Standard:

Standard:

Standard:

EPDM

FDA recommends EPDM gaskets only if medium ≤ 8% fat.

Option:

Option:

Option:

Option:

Silicone (non-Ex only)

Process connections

DIN 11850 row 2 / 11866

row A

DN2.5...150

DIN 11851

DN2.5...150

DIN 11864-2A flange with

notch

DN25...150

DIN 32676

DN25...100

ISO 2037

DN2.5...150

ISO 2852

DN2.5...150

SMS 1145

DN25...100

Tri Clamp

½

...4"

Note: DN2.5...6 (1/10...1/4") have DN10 (3/8") connections.

Electrical connections

Signal cable

Signal cable

Signal cable

Signal cable

Type A (DS)

Standard cable, double shielded.

Max. length: 600 m / 1950 ft (dep. on electrical conductivity and measuring

sensor). See documentation of the converter for more information.

Type B (BTS)

Optional cable, triple shielded.

Max. length: 600 m / 1950 ft (dep. on electrical conductivity and measuring

sensor). See documentation of the converter for more information.