TECHNICAL DATA
26
OPTIFLUX 6000
www.krohne.com
11/2013 - 4001723703 - HB OPTIFLUX 6000 R03 en
6.2 Technical data
INFORMATION!
•
The following data is provided for general applications. If you require data that is more
relevant to your specific application, please contact us or your local sales office.
•
Additional information (certificates, special tools, software,...) and complete product
documentation can be downloaded free of charge from the website (Download Center).
Measuring system
Measuring principle
Faraday's law of induction
Application range
Electrically conductive fluids
Measured value
Measured value
Measured value
Measured value
Primary measured value
Flow velocity
Secondary measured value
Volume flow
Design
Features
Hygienic design
Stainless steel housing
Food & beverage and pharmaceutical process connections
Modular construction
The measurement system consists of a flow sensor and a signal converter.
It is available as compact and as separate version.
Compact version
With IFC 050 converter: OPTIFLUX 6050 C
With IFC 100 converter: OPTIFLUX 6100 C
With IFC 300 converter: OPTIFLUX 6300 C
Remote version
In wall (W) mount version with the IFC 050 converter: OPTIFLUX 6050 W
In wall (W) mount version with the IFC 100 converter: OPTIFLUX 6100 W
In field (F), wall (W) or rack (R) mount version with IFC 300 converter:
OPTIFLUX 6300 F, W or R
Nominal diameter
DN2.5...150 / 1/10"...6"
Measuring range
-12...+12 m/s / -40...+40 ft/s
Measuring accuracy
Reference conditions
Flow conditions similar to EN 29104
Medium: Water
Temperature: +10...+30°C / +50...+86°F
Operating pressure: 1 bar / 14.5 psig
Inlet section: ≥ 5 DN
Electrical conductivity: ≥ 300 μS/cm
Wet calibrated on EN 17025 accredited calibration rig by direct volume
comparison.
Accuracy curves
Related to volume flow (MV = Measured Value)
These values are related to the pulse / frequency output.
The additional typical measuring deviation for the current output is ±10 µA.
For detailed information refer to
Measuring accuracy
on page 38.
Repeatability
±
0.1% of MV, minimum 1 mm/s
Long term stability
±
0.1% of MV
Special calibration
On request