TECHNICAL DATA

26

OPTIFLUX 6000

www.krohne.com

11/2013 - 4001723703 - HB OPTIFLUX 6000 R03 en

6.2 Technical data

INFORMATION!

The following data is provided for general applications. If you require data that is more

relevant to your specific application, please contact us or your local sales office.

Additional information (certificates, special tools, software,...) and complete product

documentation can be downloaded free of charge from the website (Download Center).

Measuring system

Measuring principle

Faraday's law of induction

Application range

Electrically conductive fluids

Measured value

Measured value

Measured value

Measured value

Primary measured value

Flow velocity

Secondary measured value

Volume flow

Design

Features

Hygienic design

Stainless steel housing

Food & beverage and pharmaceutical process connections

Modular construction

The measurement system consists of a flow sensor and a signal converter.

It is available as compact and as separate version.

Compact version

With IFC 050 converter: OPTIFLUX 6050 C

With IFC 100 converter: OPTIFLUX 6100 C

With IFC 300 converter: OPTIFLUX 6300 C

Remote version

In wall (W) mount version with the IFC 050 converter: OPTIFLUX 6050 W

In wall (W) mount version with the IFC 100 converter: OPTIFLUX 6100 W

In field (F), wall (W) or rack (R) mount version with IFC 300 converter:

OPTIFLUX 6300 F, W or R

Nominal diameter

DN2.5...150 / 1/10"...6"

Measuring range

-12...+12 m/s / -40...+40 ft/s

Measuring accuracy

Reference conditions

Flow conditions similar to EN 29104

Medium: Water

Temperature: +10...+30°C / +50...+86°F

Operating pressure: 1 bar / 14.5 psig

Inlet section: ≥ 5 DN

Electrical conductivity: ≥ 300 μS/cm

Wet calibrated on EN 17025 accredited calibration rig by direct volume

comparison.

Accuracy curves

Related to volume flow (MV = Measured Value)

These values are related to the pulse / frequency output.

The additional typical measuring deviation for the current output is ±10 µA.

For detailed information refer to

Measuring accuracy

on page 38.

Repeatability

±

0.1% of MV, minimum 1 mm/s

Long term stability

±

0.1% of MV

Special calibration

On request