Technical data – KROHNE OPTIFLUX 6000 EN User Manual
Page 27
TECHNICAL DATA
6
27
OPTIFLUX 6000
www.krohne.com
11/2013 - 4001723703 - HB OPTIFLUX 6000 R03 en
Operating conditions
Temperature
Temperature
Temperature
Temperature
Process temperature
Seperate flow sensor: -40...+140°C / -40...+284°F
Compact with IFC 300 converter: -40...+140°C / -40...+284°F
Compact with IFC 050 - IFC 100 converter: -40...+120°C / -40...+248°F
For detailed information refer to
Temperatures
on page 20.
For Ex versions different temperatures are valid.
Please check the relevant Ex documentation for details.
Ambient temperature
-40...+65°C / -40...+149°F
Storage temperature
-50...+70°C / -58...+158°F
Pressure
Pressure
Pressure
Pressure
Ambient pressure
Atmospheric
Nominal flange pressure
For detailed information refer to
Dimensions and weights
on page 30.
Vacuum load
0 mbar / 0 psi
Chemical properties
Chemical properties
Chemical properties
Chemical properties
Physical condition
Electrical Conductive liquids
Electrical conductivity
Standard: ≥ 1 μS/cm
Demineralized water: ≥ 20 μS/cm
Permissible gas content
(volume)
IFC 050: ≤ 3%
IFC 100: ≤ 3%
IFC 300: ≤ 5%
Permissible solid content
(volume)
IFC 050: ≤ 10%
IFC 100: ≤ 10%
IFC 300: ≤ 70%
Installation conditions
Installation
Assure that the flow sensor is always fully filled.
For detailed information refer to
Installation
on page 11.
Flow direction
Forward and reverse
Arrow on flow sensor indicates positive flow direction.
Inlet run
≥ 5 DN
Outlet run
≥ 2 DN
Dimensions and weights
For detailed information refer to
Dimensions and weights
on page 30.