BM102 Remote Operation - HC 275 / AMS

Dev.-Rev. EC, DD-Rev. 01

© 2000 KROHNE SA 26103 Romans (France)

Page

7

5 Status

Meaning

Meaning of General Status Messages:

Status Message

Meaning

Configuration Changed

Appears whenever a configuration parameter is changed.

More Status Available

Set when there is more information to read.
HC275 : Execute function “Watch Status”
AMS : Go to the “Status” window

Primary Variable Analog Output Fixed Set when:

Device is set to Multidrop Mode

(Polling Address

0)

Fixed Current Mode is entered

(during Loop Test and DAC Trim)

Analog Output Saturated

During normal operation the maximum value for analog output is
20mA. In case the Process Value (PV) reaches the upper sensor
limit the analog output and its digital value is saturated at 20mA
and the flag is set.

Meaning of Transmitter Specific Status Messages:

Status Message

Meaning

Invalid selection

You have chosen a non valid item

Passed parameter too large

Value too large for the parameter

Passed parameter too small

Value too small for the parameter

Too few data bytes

There are not enough bytes transmitted

In write protect mode

Protection against writing

Update failure

The BM102 is not able to measure.

Lower range value too high

The value of the range is incorrect

Applied process too high

Message appears when the PV range values are incorrect (method
Apply values)

Not in proper current mode

Appear if the current output is not fixed while we execute the
method DAC Trim

Can not change active password

Appears if the passed password is incorrect

Table non monotonous

Appears if the volume table is not monotonous

Lower range value too low

The value of the range is incorrect

Applied process too low

Message appears when the PV range values are incorrect (method
Apply values)

Upper range value too high

The value of the range is incorrect

Multidrop mode

Appears if we are in multidrop mode and we want to fixe a value
on the current output

Illegal password symbol

The characters of the passed password are incorrect. They must be
‘E’, ‘R’ or ‘U’

Upper range value too low

The value of the range is incorrect

Invalid units code

The unit is not supported by the BM102

Invalid function

The result of the function is incorrect.

Invalid password

Can not enter Specialist menu because the password is incorrect.

Access restricted

When the customer password is set, this message appears if we
want to modify a parameter.

Invalid range units code

Appears if the unit of the range is incorrect

Device busy

Appears when the device is busy. Normally it appears when the