beautypg.com

Capaflux ifm 5080 k-cap electromagnetic flowmeter – KROHNE CAPAFLUX IFM 5080 K Datasheet EN User Manual

Page 2

background image

Capaflux IFM 5080 K-CAP
Electromagnetic Flowmeter

... non-contact process flow measurment
from 0.05 µS/cm electrical conductivity

non-contact flow measurement

no electrodes

easy to specify

unimpeded flow cross-sectional area

optimum flow shaping

resistant to abrasion

ceramic measuring tube

dimensionally stable vacuum-resistant

outstanding accuracy

Special advantages

capacitive electrodes for non-contact measurement.

the measuring section is resistant to abrasion from even high solids

contents.

the ceramic measuring tube is dimensionally stable and vacuum-

resistant.

the special shape of the measuring tube helps to optimize the flow

profile, even with minimum pressure drop, refer to diagram on

page 3.

the measuring error is less than 0.5% of the measured value.

the integral design ensures easy installation, safe and reliable

operation.

the crevice-free measuring tube has no blind spots and conforms

to food requirements, the ceramic surface is ultrasmooth,

R

a

< 0,8 µm surface finish.

ATEX approval

Ex II 2 GD KEMA 01 ATEX 2232X

CAPAFLUX IFM 5080 K/CAP-EEx:

EEx d IIC T6 … T4

EEx de IIC T6 … T4

CAPAFLUX IFM 5080 K/CAP/i-EEx:

with outputs intrinsic safety

EEx d [ia] IIC T6 … T4

EEx de [ia] IIC T6 … T4

2

No restrictions ...

... through insulating products with a

film-forming tendency:

asphalt, latex suspensions

... through low electrical conductivity:

ultrahigh-purity water, alcohols,

glycerins, glycols

... through high solids contens:

fruit pieces, pulps, concrete

... for sterile processes:

chemical and food industries

... when used in hazardous areas:

ATEX approval

FM certification pending

... through electrode materials:

the capacitive electrodes are located

behind the ceramic tube, i.e. non-

contact measurement, no contact with

the process product.

Calibrated on EN 17 025

accredited calibration rigs,

accuracy of calibration better

than 99.97% of the measured value.