Operation – KROHNE OPTISENS PH 9100 EN User Manual
Page 30

4
OPERATION
30
OPTISENS PH 9100
www.krohne.com
09/2012 - 4002314801 - MA OPTISENS PH 9100 R02 en
• Enter the measured temperature. Press ^ to confirm the entered value. The manually
measured temperature will now be used for temperature compensation.
• Step 2c: probe process input A
The temperature sensor connected to input A of the signal converter is used for temperature
compensation.
The temperature sensor probe process input A has to configured in submenu process input A.
If you use a Pt100/1000 follow the instruction in step 2.1. If you use manually entered value,
follow the instruction in step 2.2.
Step 2b: probe manual
Measure the temperature of the measuring medium.
•
Press > for more than 2.5 seconds, then release the button. You are on the main menu level. In the
upper line of the display "A" appears, beneath the main menu setup is highlighted.
•
Press or until the main menu setup is highlighted.
MAIN MENU
A quick setup
B test
> C setup
D service
•
Press > to enter the chosen menu.
You are on the first submenu level. In the upper line of the display "setup" and "c1" appears,
beneath the submenu process input A is highlighted. Choose process input A or B where ever
pH is configurated.
•
Press or to select process input A or process input B.
•
Press > to enter the chosen menu process input A/B.
You are on the second submenu level. Press or until the submenu temperature is
highlighted.
•
Press > to enter the chosen menu. The submenu probe is highlighted.
•
Press or until the submenu manual is highlighted.
•
Press ^ to enter the chosen menu.