Installation and Operating Instructions TTP x00

4 Technical

data

4.1 Technical

data

Connection head

Stainless steel V2A 1.4305
Cable gland M16 (only TTP 200/400)

Electrical connection

M12 connector

Type of protection

IP 67

Ambient temperature

–20...+60 °C
Threaded socket G1/2“h; V4A 1.4571
Threaded socket M12h; V4A 1.4571
Push-in connection via protective tube dia. 6 mm

Process connections

Threaded socket G1/2; V4A 1.4571

Protective tube

Dia. 6 mm x 1 mm; V4A 1.4571

Positioned lengths (mm)

20, 50, 100, 150, 250 (others on request)

Sensor element

Pt100 DIN EN 60751 Cl. A (dual type also possible)

Response time t90

Dia. 6 mm: 7 s; dia. 4 mm: 6 s; dia. 3 mm: 1.5 s

Process temperature range

–20...+170 °C

Operating pressure

Max. 16 bar

With direct Pt100 connection
Output

Screw terminals for 2-, 3-, 4-wire connection

With integrated head transmitter
Supply voltage

8...28 V DC (from current loop)

Output 4...20

mA

Accuracy

± 0.1% of full-scale range