KROHNE TTP 400 EN User Manual
Page 7
Installation and Operating Instructions TTP x00
7
3
Fault diagnosis and corrective action
In the event of a fault or maloperation, please go through the various faults listed in the table. Do
not attempt to disassemble the device.
Should you fail to locate the fault, please contact our Technical Service.
Fault Cause Action/elimination
With Pt100 connection
Incorrect display
Terminal assignment is
incorrect
Check terminal assignment
Cable break
Measure continuity
Error display
Short-circuit Check
wiring
With integrated signal converter
Supply voltage at module
< 8 V
Load impedance too high,
voltage too low
Incorrect current signal
Incorrect measuring range
Correct the programming
Cable break
Check continuity of cables
No current signal
Supply polarity incorrect
Reverse terminals 1, 2
Current signal > 23 mA
Short-circuit
Check wiring