14

Installation and Operating Instructions TTP x00


6.3.2 Pt100 tips and response times

Tube 6 mm (S 6)

Tube 4 mm (S 4)

Tube 3 mm (S 3)

Halftime: T

50

< 3.0 sec.

Halftime:

T

50

< 2.5 sec.

Halftime:

T

50

< 0.6 sec.

90% - time: T

90

< 7.0 sec.

90% - time:

T

90

< 6.0 sec.

90% - time:

T

90

< 1.5 sec.





6.4 Features


• Sensors for hygienic installation, no elastomers

• Compact design, optionally with head transmitter

• For direct connection to an SPC or PLC (with integrated head transmitter)

• High accuracy (DIN EN 60751 Class A)

• Very fast response times

• Optimised flow geometry

• Food-compatible materials

• No maintenance requirement

• Installation in pipelines DN 15 and higher

• Defined position of the cable connection

• Hygienic adapter sleeves for other process connections

Tube

Pt100 Chip

This manual is related to the following products: