6 Set up

6.1 General

The numbers in brackets refer to the following illustrations.

The switching status of the electronics can be checked when

the cover of the plastic housing is closed (signal lamp). In the

basic setting, products with density > 0.7 g/cm³ (0.025 lbs/in³)

can be detected. For products with lower density, the switch

must be set to > 0.5 g/cm³ (0.018 lbs/in³).
On the electronics module you will find the following indicating

and adjustment elements:

l

Signal lamp (1)

l

DIL switch for sensitivity adjustment (2)

Mode switch (A/B)

On the signal conditioning instrument via the signal con-

ditioning instrument (only permissible in mode A when used

acc. to WHG).
The switching condition can be changed with the A/B switch.

You can set the required mode acc. to the "Function chart" (A -

max. detection or overfill protection, B - min. detection or dry

run protection). The integration time can also be modified on

the signal conditioning instrument.

Note:

For test purposes, immerse the tuning fork of OPTISWITCH

5200 C, 5250 C always in liquids. Do not test the function of

OPTISWITCH 5200 C, 5250 C with the hand. This can

damage the sensor.

Function/Configuration

20

OPTISWITCH 5200 C, 5250 C - with two-wire output

Set up