7 Tighten the terminal screws (3) alternately. Make sure that

the gap between clamp and pressure screw is in parallel.

Tighten the terminal screws up to a torque of 10 ±1 Nm

(7 ±0.7 lbf ft).
Hence the clamp (1) is pressed against the tube and fixes

the tube of the sensor in this position

8 Fasten the locking angle (4) with the hold screw (6) and the

spring ring (5) laterally on the pressure screw (11)
Hence the pressure screw and the terminal screws are

protected against unintentional loosening

Lock fitting ARV 67 - up to 16 bar for OPTISWITCH 3300 C

7

Mounting

30925

-EN