beautypg.com

Electrical connections – KROHNE OPTIFLEX 4300C ATEX EN User Manual

Page 14

background image

3

ELECTRICAL CONNECTIONS

14

OPTIFLEX 4300 C

www.krohne.com

02/2009 - 4000523601 - AD ATEX OPTIFLEX 4300 R01 en

3.5.2 Maximum intrinsically-safe values for the electrical circuit

Level transmitter without external sensor (output terminal 1 or 2)

• U

i

≤30 V

• I

i

≤300 mA

• P

i

≤1 W

• C

i

=30 nF

• L

i

=0.2 mH

Level transmitter with external sensor: output terminal 1

• U

i

≤30 V

• I

i

≤300 mA

• P

i

≤1 W

• C

i

=30 nF

• L

i

=0.2 mH

Level transmitter with external sensor: output terminal 2

• C

i

=0 nF

• L

i

=0 mH

3.5.3 Supply voltage

Level transmitter without external sensor

Level transmitter with external sensor

Current output terminal

Minimum voltage at output

terminal [VDC]

Maximum voltage at output

terminal [VDC]

1 (U

S1

)

14

1

30

1

2 (U

S2

)

2

10

1

30

1

1 for a current output of 22 mA
2 use a second power supply. This power supply energizes current output terminal 2 only.

Current output terminal

Minimum voltage at output

terminal [VDC]

Maximum voltage at output

terminal [VDC]

1 (U

S1

)

14

1

30

1

2 (U

S2

)

2

3

3

1 for a current output of 22 mA
2 use a second power supply. This power supply energizes current output terminal 2 only.
3 for more data, refer to the data sheet for the external sensor

AD_ATEX_OPTIFLEX4300_en_090528_4000523601_R01_D5D.book Page 14 Thursday, May 28, 2009 6:00 PM