2

INSTALLATION

12

OPTIFLEX 1300 C

www.krohne.com

06/2010 - 4000091403 - AD ATEX OPTIFLEX 1300 R05 en

Equipment category 2 G, 2/3 G or 3 G:
Ex ia and Ex d[ia] devices - with the PROFIBUS PA or FOUNDATION™ Fieldbus output options

Equipment category 1 D: Ex iaD devices only
Equipment category 1/2 D, 2 D, 2/3 D or 3 D: Ex iaD and Ex tD[iaD] devices

Temperature

class

Ambient temperature

Flange temperature

Ø2 mm / Ø0.08¨ single

cable probe (HT

version)

Ø2 mm / Ø0.08¨ single

cable probe (other

versions)

All other probe types

[°C]

[°F]

[°C]

[°F]

[°C]

[°F]

[°C]

[°F]

T6

-40…+51

-40…+124

-40…+50

-40…+134

-40…+51

-40…+124

-50…+60

1

-58…+140

1

-40…+48

-40…+118

-40…+44

-40…+111

-40…+47

-40…+116

-50…+85

1

-58…+185

1

T5

-40…+72

-40…+161

-40…+72

-40…+161

-40…+72

-40…+161

-50…+75

1

-58…+167

1

-40…+68

-40…+154

-40…+64

-40…+147

-40…+67

-40…+152

-50…+100

1

-58…+212

1

T4

-40…+80

-40…+176

-40…+80

-40…+176

-40…+80

-40…+176

-50…+85

1

-58…+185

1

-40…+75

-40…+167

-40…+72

-40…+161

-40…+75

-40…+167

-50…+110

1

-58…+230

1

-40…+72

-40…+161

-40…+65

-40…+149

-40…+70

-40…+158

-50…+135

1

-58…+275

1

T3

-40…+70

-40…+158

-40…+60

-40…+140

-40…+65

-40…+149

-50…+150

1

-58…+302

1

-40…+66

2

-40…+160

2

-

-

-40…+60

2

-40…+140

2

-50…+180

1

-58…+356

1

-40…+64

2

-40…+151

2

-

-

-40…+55

2

-40…+131

2

-50…+200

1

-58…+392

1

T2

-40…+57

2

-40…+134

2

-

-

-

-

-50…+250

1

-58…+482

1

-40…+50

2

-40…+122

2

-

-

-

-

-50...+300

1

-58…+572

1

1 Min. -50°C / -58°F, if an EPDM gasket is used. Min. -40°C / -40°F, if an FKM/FPM gasket is used. Min. -20°C / -4°F, if a Kalrez® 6375

gasket is used.

2 If the device has an adaptor for BM 100 probes, the flange temperature must be less than +150°C / +302°F

Ambient temperature

Flange temperature

Ø2 mm / Ø0.08¨ single

cable probe (HT version)

Ø2 mm / Ø0.08¨ single

cable probe (other

versions)

Other probes

[°C]

[°F]

[°C]

[°F]

[°C]

[°F]

[°C]

[°F]

-40…+85

-40…+185

-50…+300

1

-58…+572

1

-50…+150

1

-58…+302

1

-50…+200

1

-58…+392

1

1 Min. -50°C / -58°F, if an EPDM gasket is used. Min. -40°C / -40°F, if an FKM/FPM gasket is used. Min. -20°C / -4°F, if a Kalrez® 6375

gasket is used.