Installation – KROHNE OPTIFLEX 1300C ATEX EN User Manual
Page 10
2
INSTALLATION
10
OPTIFLEX 1300 C
www.krohne.com
06/2010 - 4000091403 - AD ATEX OPTIFLEX 1300 R05 en
Equipment category 1 G:
Ex ia devices only - with the 4...20 mA output option
Equipment category 1 G:
Ex ia devices only - with PROFIBUS PA or FOUNDATION™ Fieldbus output options
Equipment category 1/2 G:
Ex ia and Ex d[ia] devices - with the 4...20 mA output option
Equipment category 1/2 G:
Ex ia and Ex d[ia] devices - with the PROFIBUS PA or FOUNDATION™ Fieldbus output options
Temperature
class
Ambient temperature
Flange temperature
[°C]
[°F]
[°C]
[°F]
T6
-20…+57
-4…+134
-20…+60
-4…+140
Temperature
class
Ambient temperature
Flange temperature
Ø2 mm / Ø0.08¨ single
cable probe (HT
version)
Ø2 mm / Ø0.08¨ single
cable probe (other
versions)
All other probe types
[°C]
[°F]
[°C]
[°F]
[°C]
[°F]
[°C]
[°F]
T6
-20…+51
-4…+124
-20…+50
-4…+122
-20…+51
-4…+124
-20…+60
-4…+140
Temperature
class
Ambient temperature
Flange temperature
[°C]
[°F]
[°C]
[°F]
T6
-40…+57
-40…+134
-20…+60
-4…+140
T5
-40…+72
-40…+161
-20…+60
-4…+140
T4
-40…+80
-40…+176
-20…+60
-4…+140
Temperature
class
Ambient temperature
Flange temperature
Ø2 mm / Ø0.08¨ single
cable probe (HT
version)
Ø2 mm / Ø0.08¨ single
cable probe (other
versions)
All other probe types
[°C]
[°F]
[°C]
[°F]
[°C]
[°F]
[°C]
[°F]
T6
-40…+51
-40…+124
-40…+44
-40…+111
-40…+51
-40…+124
-20…+60
-4…+140
T5
-40…+72
-40…+161
-40…+72
-40…+161
-40…+72
-40…+161
-20…+60
-4…+140
T4
-40…+80
-40…+176
-40…+80
-40…+176
-40…+80
-40…+176
-20…+60
-4…+140