beautypg.com

KROHNE OPTISOUND 3030C 2wire EN User Manual

Page 41

background image

Materials
- Housing

ABS

- Inspection window

Polyester foil

Supply voltage
Supply voltage

14 … 36 V DC

Permissible residual ripple
- <100 Hz

U

ss

<1 V

- 100 Hz … 10 kHz

U

ss

<10 mV

Load

see diagram

1000

750

500

250

14

18

16

20

22

24

26

28

30

32

34

36

Ω

V

4

1

2

3

Fig. 26: Voltage diagram

1

HART load

2

meaningless

3

Voltage limit

4

Supply voltage

Electrical protective measures
Protection
- Plastic housing

IP 66/IP 67

- Alu and stainless steel housing

IP 66/IP 68 (0.2 bar)

5)

Overvoltage category

III

Protection class

II

5)

Requirement to maintain the protection is the suitable cable.

OPTISOUND 3030 C - 4 … 20 mA/HART

41

Supplement

30511

-EN

-050914