Zero-point measurement

For ultrasonic flowmeters it is important to have a good zero-point. For making a zero-point it is very important to have a situation where the flow
is absolutely zero. If that is not possible the default zero-point can be used, which is a good approximation.

Have the fluid flow at a high flow rate for at least five minutes. Stop the flow and shut down the upstream and downstream valves (if available).

Let the fluid stabilize until the displayed flow rate becomes stable. (At least a few minutes, but even longer for large diameters (> DN 300).

Make a zero-point measurement (Fct. 3.1.4.) and check if the display comes to zero. Else repeat the zero-point measurement until a stable
zero-point is obtained.

Now the UFM 800W ultrasonic flowmeter should be operating in correct way.