beautypg.com

Terms and definitions, 1 description of the considered profiles – KROHNE DK 32-DK 34-DK 37 Safety V2 EN User Manual

Page 4

background image

2

TERMS AND DEFINITIONS

4

DK32 - DK34 - DK37

www.krohne.com

06/2013 - 4000574703 MA DK32 34 37 SIL R03 en

Terms and definitions

Terms and definitions

2.1 Description of the considered profiles

DC

D

Diagnostic Coverage of dangerous failures

FIT

Failure In Time (1x10

-9

failures per hour)

FMEDA

Failure Modes, Effects and Diagnostic Analysis

HFT

Hardware Fault Tolerance

Low demand mode Mode, where the frequency of demand for operation made on a safety-related

system is not greater than one per year and not greater than twice in the proof test

frequency.

PFD

AVG

Average Probability of Failure on Demand

SFF

Safe Failure Fraction summarizes the fraction of failure, which lead to a safe state

and the fraction of failures which will be detected by diagnostic measures and lead

to a defined safety action.

SIF

Safety Instrumented Function

SIL

Safety Integrity Level

Type A component

"Non-complex" subsystem (all failure modes are well defined); for details see

7.4.3.1.2 of IEC 61508-2.

Type B component

"Complex" subsystem (all failure modes are well defined); for details see 7.4.3.1.2

of IEC 61508-2.

T[Proof]

Proof Test Interval

Profile

Profile according to
IEC 60654-1

Ambient temperature [°C]

Temperature cycle
[°C / 365 days]

Average (external)

Mean (inside box)

2

C3

25

30

25

4

D1

25

30

35

Profile 2 (low stress):

Mechanical field products with minimal self heating, subject to daily
temperature swings.

Profile 4 (high stress):

Unprotected mechanical field products with minimal self heating,
subject to daily temperature swings and rain or condensation.

MA_DK32_34_37_SIL_R03_en_574703_PRT.book Page 4 Wednesday, June 26, 2013 9:17 AM