beautypg.com

Electrical connections, 4 earthing and equipotential bonding – KROHNE H250 M10 ATEX II3D Ex t EN User Manual

Page 11

background image

ELECTRICAL CONNECTIONS

4

11

H250 M10

www.krohne.com

03/2011 - 4001348101- MA H250/M10-Ex-II3G-AD R01 en

4.4 Earthing and equipotential bonding

The signal converter shall be connected to the equipotential bonding system of the hazardous
area via the external grounding connection on the signal converter housing. The measuring unit
and the signal converter are electrically connected via an equipotential bonding conductor.

Any existing cable shields should be connected to earth according to applicable installation
regulations (EN 60079-14). A terminal connection in the terminal compartment permits a short
way earthing of the cable shields.

1 External grounding connection
2 Internal grounding connection

AD_H250_M10_II3D_R01_en_4001248101_PRT.book Page 11 Wednesday, March 30, 2011 11:44 AM